ATI 100X Filter Tester
The 100X Automated Filter Tester is designed for test and quality control validation of filter media, cartridges and masks used in medical and industrial hygiene applications. The 100X meets leading global industry standards such as NIOSH 42 CFR Part 84, GB 2626 and EN 13274-7:2019. The 100X is a high capacity automated solution ideally suited to meet the precise and rigorous requirements of both development and production quality control environments. The 100X is perfect for testing N95, KN95 and FFP respirator masks that are widely used to help prevent the spread of the new Coronavirus.
| Performance & Product Specifications | |||
| Unit | 100Xp (Oil) | 100Xp (Paraffin) | 100Xs (Salt) |
| Standards | NIOSH 42CFR Part 84 or Equivalent | EN13274-7:2019 | NIOSH 42CFR Part 84 or Equivalent |
| Reagents | DOP, PAO-4 | Paraffin Oil | NaCl |
| Primary Aerosol Reagent | CAS# 8012-95-1 Density: 0.818 – 0.875 g/cm³ Dynamic Viscosity: 0.025Pa*s – 0.080 Pa*s Kinematic Vicosity: <35mm²/s Kinematic Viscosity: @40C: 13.5-16.5mm²/s |
4% NaCl solution by weight | |
| Replenishment Reagent | 0.9% NaCl solution by weight | ||
| Available Configurations | Manual, Hybrid and Automatic | ||
| Aerosol Generation | |||
| Particle Size | |||
| Count Median Diameter (CMD) | 0.185μm ± 0.02 | 0.29 – 0.45μm | 0.075 μm ± 0.02 |
| Mass Mean Diameter (MMD) | 0.3μm | Not specified by the Standard | 0.26 μm |
| Geometric Standard Deviation (GSD) | < 1.6 | 1.6 – 2.2 | < 1.86 |
| Concentration (mg/m3) | 80 – 120 | 15 – 25 | 15 – 25 |
| Concentration (mg/m3) High Flow | 40 – 60 | ||
| Flow Rate Options | 10 – 120 L/min (0.18 – 4.2 SCFM) | ||
| 10 – 40 L/min (0.18 – 1.4 SCFM) | |||
| High Flow Option | 10 – 170 L/min (0.18 – 6.0 SCFM) |
||
| Aerosol Flow Accuracy | Accurate to ± 0.4% of reading, plus +/- 0.2% full scale | ||
| Aerosol Detection | |||
| Dynamic Range | 0.1 μg/m³ to >200 mg/m³, accurate to ± 1% of reading | ||
| Penetration | 0.0005% | ||
| Efficiency | 99.9995% | ||
| Pressure Management | |||
| Pressure Range Options | 0-25 mm H²O/0-245Pa | ||
| 0-50 mm H²O/0-490Pa | |||
| 0-100 mm H²O/0-981Pa | |||
| 0-250 mm H²O/0-2452Pa | |||
| Accuracy | ±0.25% of full scale | ||
| Test Fixture | Standard Size: ID 11.4 cm (4.5”) | ||
| Stroke Height: 14.6 cm (5.75”) | |||
| Custom Test Fixture | On Unit: ≤ 20.3 cm (8″) OD | ||
| Off Unit: No Size Limit | |||
| Communication Ports | USB; RS232; Data Connection; PLC-Controls (Automatic, Hybrid version) | ||
| Utilities | |||
| Power | 110VAC , 5.5A; or 220VAC, 2.5A (50/60 Hz) | ||
| Compressed Air | 396 L/min at 6.2 bars +- 0.35 Bar (620 KPa +- 35 KPa) or 14 SCFM at 90 psig +- 5.0 psig |
||
| Air Supply Line ID | >9.5 mm (0.375”) | ||
| Dimensions (L x W x H) | |||
| 100X Main Unit | 71 cm x 69 cm x 89 cm (28” x 27” x 35”) | ||
| Optional Base | 71 cm x 69 cm x 76 cm (27” x 28” x 30”) | ||
| Optional Exhaust | 42.5 cm x 62.8 cm x 40.6 cm (16.75″ x 24.75″ x 16″) | ||
| Weight | |||
| 100X Main Unit | 75 kg (165 lbs) | ||
| Optional Base | 38.6 kg (85 lbs) | ||
| Optional Local Exhaust | 20 kg (44 lbs) | ||
| Compliance | CE, RoHS, FCC, CSA | ||